Improving Real-Time Feedback during Cochlear Implantation: The Auditory Nerve Neurophonic/Cochlear Microphonic Ratio

Title: Improving Real-Time Feedback during Cochlear Implantation: The Auditory Nerve Neurophonic/Cochlear Microphonic Ratio
Authors: Bester, C; O'Leary, SJ; Venail, F; Büchner, A; Félix, TR; Lai, WK; Boccio, C; Choudhury, B; Tejani, V; Mylanus, E; Sicard, M; Lenarz, T; Birman, C; Di Lella, F; Roland, JT; Gantz, B; Plant, K; English, R; Arts, R; Beynon, A
Publisher Information: Ovid Technologies (Wolters Kluwer Health)
Publication Year: 2025
Collection: The University of Melbourne: Digital Repository
Description: OBJECTIVES: Real-time monitoring of cochlear function to predict the loss of residual hearing after cochlear implantation is now possible. Current approaches monitor the cochlear microphonic (CM) during implantation from the electrode at the tip of the implant. A drop in CM response of >30% is associated with poorer hearing outcomes. However, there is prior evidence that CM amplitude can fluctuate in a manner unrelated to hearing trauma, leading to false positives. By monitoring another cochlear response, the auditory nerve neurophonic (ANN), a differentiation between CM drops that result in reduced cochlear output from false positives may be possible. The hypothesis tested in the present work was that ANN/CM ratios measured during a CM drop will increase during drops not associated with postoperative hearing loss. DESIGN: Twenty-eight adult participants with known CM drops during implantation were taken from a larger data set. This contains adult cochlear implant candidates scheduled to receive a Cochlear Nucleus cochlear implant with either the slim-straight or slim-modiolar electrode array with preoperative audiometric low-frequency average thresholds of ≤80 dB HL at 500, 750, and 1000 Hz in the ear to be implanted. Patients were recruited from eight international implant sites. Pure-tone audiometry was measured postoperatively and 4 to 6 weeks after implantation. Electrocochleography was measured during and immediately after the implantation of the array in response to a 500-Hz, 6-msec pure-tone pip at 110 dB HL. RESULTS: The ANN/CM ratio rose during CM drops in 19 of these patients and decreased in 9. At the follow-up timepoint, patients with a decreasing ANN/CM ratio had a median hearing loss of 29.0 dB, significantly worse than the group with increasing ratio at 13.3 dB ( p = 0.004). Considering only the change in ANN amplitude during a CM drop led to smaller groups (ANN drop during CM drop N = 17, ANN increasing during CM drop N = 6) due to 5 patients having undetectable ANN during the CM drop. Using ...
Document Type: article in journal/newspaper
Language: English
ISSN: 0196-0202
Relation: https://hdl.handle.net/11343/359761
Availability: https://hdl.handle.net/11343/359761
Rights: https://creativecommons.org/licenses/by-nc-nd/4.0 ; CC BY-NC-ND
Accession Number: edsbas.34C69583
Database: BASE