| Title: |
The effect of cooling rate on the precision of measurement of glow peak 5 in the thermoluminescence of LiF:Mg,Ti (TLD-100) |
| Authors: |
Bokobza, Yogev; Biderman, Shlomo; Horowitz, Yigal Shalom; Oster, Leonid; Eliyahu, Ilan; Shapiro, Alex; Nemirovsky, Dmitry; Reshes, Galina; Herman, Beny; Einav, Hezy |
| Source: |
Radiation Protection Dosimetry ; volume 200, issue 4, page 423-427 ; ISSN 0144-8420 1742-3406 |
| Publisher Information: |
Oxford University Press (OUP) |
| Publication Year: |
2024 |
| Description: |
The effect of natural rapid cooling and oven slow cooling on the precision of thermoluminescence measurements of LiF:Mg,Ti is investigated. Three separate series of measurements resulted in average precisions of 5.1 and 5.0%, respectively. However, the highest precision of 1.7% (1 SD) was achieved for an oven-cooled material. |
| Document Type: |
article in journal/newspaper |
| Language: |
English |
| DOI: |
10.1093/rpd/ncad318 |
| Availability: |
https://doi.org/10.1093/rpd/ncad318; https://academic.oup.com/rpd/article-pdf/200/4/423/57027570/ncad318.pdf |
| Rights: |
https://academic.oup.com/pages/standard-publication-reuse-rights |
| Accession Number: |
edsbas.76E7074 |
| Database: |
BASE |