The risk of failure of subcutaneous implantable cardioverter defibrillator therapy: from PRAETORIAN score to clinical practice

Title: The risk of failure of subcutaneous implantable cardioverter defibrillator therapy: from PRAETORIAN score to clinical practice
Authors: Ziacchi, Matteo; Ottaviano, Luca; Checchi, Luca; Viani, Stefano; Nigro, Gerardo; Bianchi, Valter; De Bonis, Silvana; De Filippo, Paolo; Francia, Pietro; Rapacciuolo, Antonio; Vitulano, Gennaro; Perego, Giovanni Battista; Schillaci, Vincenzo; Lavalle, Carlo; Migliore, Federico; Pisanò, Ennio C L; Compagnucci, Paolo; Palmisano, Pietro; Botto, Gianluca; Rordorf, Roberto; Lovecchio, Mariolina; Valsecchi, Sergio; Biffi, Mauro
Contributors: Ziacchi, Matteo; Ottaviano, Luca; Checchi, Luca; Viani, Stefano; Nigro, Gerardo; Bianchi, Valter; De Bonis, Silvana; De Filippo, Paolo; Francia, Pietro; Rapacciuolo, Antonio; Vitulano, Gennaro; Perego, Giovanni Battista; Schillaci, Vincenzo; Lavalle, Carlo; Migliore, Federico; Pisanò, Ennio C L; Compagnucci, Paolo; Palmisano, Pietro; Botto, Gianluca; Rordorf, Roberto; Lovecchio, Mariolina; Valsecchi, Sergio; Biffi, Mauro
Publisher Information: Oxford; Oxford University Press
Publication Year: 2025
Collection: Sapienza Università di Roma: CINECA IRIS
Subject Terms: conversion; defibrillation test; implantable defibrillator; subcutaneous
Description: Background: The subcutaneous implantable cardioverter-defibrillator (S-ICD) is an alternative to traditional ICDs. The PRAETORIAN score, based on chest radiographs, has been validated to predict the probability of successful S-ICD defibrillation testing by assessing factors like fat thickness between the coil and sternum and generator placement. Objective: This study evaluated the correlation between the PRAETORIAN score and clinical characteristics, as well as implantation variables. Methods: We retrospectively analyzed data from 1253 patients who had undergone implantation of an S-ICD across 33 centers. The intermuscular positioning of the pulse generator was adopted in all patients. Post-implantation posterior-anterior and lateral chest radiographs were analyzed to calculate the PRAETORIAN score. Results: 95.7% of patients had a PRAETORIAN score
Document Type: article in journal/newspaper
Language: English
Relation: info:eu-repo/semantics/altIdentifier/pmid/39834232; info:eu-repo/semantics/altIdentifier/wos/WOS:001421116300001; volume:27; issue:2; numberofpages:8; journal:EUROPACE; https://hdl.handle.net/11573/1732276
DOI: 10.1093/europace/euaf011
Availability: https://hdl.handle.net/11573/1732276; https://doi.org/10.1093/europace/euaf011
Rights: info:eu-repo/semantics/openAccess